A1R-A1 Specifications
A1R
| Input/output port: |
Laser input port: 3 (FC x2, direct x1) Signal output port: 4 (SMA x2, FC x1, VAAS x1) |
|---|---|
| Laser |
|
| Wavelength and power: |
405LD: max. 38mW, Multi-Ar (457/488/514): max. 65mW, 488DPSS: max. 75mW, 561DPSS: max. 25mW, 543HeNe: max. 1mW, 638LD: max. 10mW 440LD: max. 15mW (available as option) |
| Modulation: |
Method: AO (Acousto) device or drive current control Control: power control for each wavelength, Return mask, ROI exposure control |
| Laser unit: |
Standard: LU4 4-laser unit Optional: C-LU3EX 3-laser unit EX |
| Standard fluorescence detector |
|
| Detector: |
4 PMT |
| Filter cube: |
6 filter cubes commonly used for a microscope mountable on each of three filter wheels Recommended wavelengths: 450/50, 482/35, 525/25, 595/50, 700/75, 540/30, 515/30, 585/65 |
| Diascopic detector |
|
| Detector: |
PMT |
| Scanning head |
|
| Scanning: |
Scanning range: square inscribed in a ø18mm circle Standard image acquisition Scanner: non-resonant scanner x2 Pixel size: max. 4096 x 4096 pixels Scanning speed: 4 fps (512 x 512 pixels) Zoom: 1-1000x continuously variable Scanning mode: X-Y, XY rotation, Free line, Line Z High-speed image acquisition Scanner: resonant scanner (X-axis, resonance frequency 7.8kHz), non-resonant scanner (Y-axis) Pixel size: max. 512 x 512 pixels Scanning speed: 30 fps (512 x 512 pixels) to 230 fps (512 x 64 pixels), 15,600 lines/sec (line speed) Zoom: 7 steps (1x, 1.5x, 2x, 3x, 4x, 6x, 8x) Scanning mode: X-Y, Line Acquisition method: Standard image acquisition, High-speed image acquisition, Simultaneous photo activation and image acquisition |
| Dichroic mirror: |
Low-angle incidence method Position: 8 Standard filter: 405/488, 405/488/561, 405/488/561/638, 405/488/543/638, 457/514, BS20/80 |
| Pinhole: |
12-256µm variable (1st image plane) |
| Spectral detector |
|
| Number of channels: |
32 channels |
| Spectral image acquisition speed: |
4 fps (256 x 256 pixels), 1000 lps |
| Maximum wavelength and resolution: |
80nm (2.5nm), 192nm (6nm), 320nm (10nm) Wavelength range variable in 0.25nm steps |
| Unmixing: |
High-speed unmixing, Precision unmixing |
| Z step: |
0.025µm |
| Compatible microscopes: |
ECLIPSE Ti-E inverted microscope, ECLIPSE TE2000-E inverted microscope, ECLIPSE 90i upright microscope, ECLIPSE FN1 fixed stage microscope |
| Option: |
Motorized XY stage, High-speed Z stage, VAAS |
| Software |
|
| Display/image generation: |
2D analysis, 3D volume rendering/orthogonal, 4D analysis, spectral unmixing |
| Image format: |
JP2, JPG, TIFF, BMP, GIF, PNG, ND2, JFF, JTF, AVI, ICS/IDS |
| Application: |
FRAP, FLIP, FRET, photo activation, three-dimensional time-lapse imaging, multipoint time-lapse imaging, colocalization |
| Control computer |
|
| OS: |
Microsoft Windows® XP 32bit SP (English version) |
| CPU: |
Intel Xeon 5160 (3GHz/1333MHz/dual core) or higher |
| Memory: |
4GB or more |
| Hard disk: |
SAS (15,000rpm), 160GB or more x2, RAID 0 configuration |
| Data transfer: |
Dedicated data transfer I/F |
| Monitor: |
1600 x 1200 or higher resolution, 2 LCD monitor configuration recommended |
| Installation condition: |
Temperature 5 - 35°C, humidity 65% (RH) or less (non-condensing) |
| Power Source |
|
| Controller: |
Input voltage: 100–240VAC ±10% 50–60Hz Current rating: 5A @100VAC Overcurrent protection: main breaker 15A |
| LU-LR 4-laser power source rack: |
Power source for Ar laser and control circuit: 100VAC, 15A/115VAC, 15A/230VAC, 7.5A, 50/60Hz (breaker 15A) Power source for lasers except Ar laser: 100VAC, 3A/115VAC, 3A/230VAC, 1.5A, 50/60Hz (breaker 5A) |
A1
| Input/output port: |
Laser input port: 3 (FC x2, direct x1) Signal output port: 4 (SMA x2, FC x1, VAAS x1) |
|---|---|
| Laser |
|
| Wavelength and power: |
405LD: max. 38mW, Multi-Ar (457/488/514): max. 65mW, 488DPSS: max. 75mW, 561DPSS: max. 25mW, 543HeNe: max. 1mW, 638LD: max. 10mW 440LD: max. 15mW (available as option) |
| Modulation: |
Method: AO (Acousto) device or drive current control Control: power control for each wavelength, Return mask, ROI exposure control |
| Laser unit: |
Standard: LU4 4-laser unit Optional: C-LU3EX 3-laser unit EX |
| Standard fluorescence detector |
|
| Detector: |
4 PMT |
| Filter cube: |
6 filter cubes commonly used for a microscope mountable on each of three filter wheels Recommended wavelengths: 450/50, 482/35, 525/25, 595/50, 700/75, 540/30, 515/30, 585/65 |
| Diascopic detector |
|
| Detector: |
PMT |
| Scanning head |
|
| Scanning: |
Standard image acquisition Scanner: non-resonant scanner x2 Pixel size: max. 4096 x 4096 pixels Scanning speed: 4 fps (512 x 512 pixels) Zoom: 1-1000x continuously variable Scanning mode: X-Y, XY rotation, Free line, Line Z |
| Dichroic mirror: |
Low-angle incidence method Position: 8 Standard filter: 405/488, 405/488/561, 405/488/561/638, 405/488/543/638, 457/514, BS20/80 |
| Pinhole: |
12-256µm variable (1st image plane) |
| Spectral detector |
|
| Number of channels: |
32 channels |
| Spectral image acquisition speed: |
4 fps (256 x 256 pixels), 1000 lps |
| Maximum wavelength and resolution: |
80nm (2.5nm), 192nm (6nm), 320nm (10nm) Wavelength range variable in 0.25nm steps |
| Unmixing: |
High-speed unmixing, Precision unmixing |
| Z step: |
0.025µm |
| Compatible microscopes: |
ECLIPSE Ti-E inverted microscope, ECLIPSE TE2000-E inverted microscope, ECLIPSE 90i upright microscope, ECLIPSE FN1 fixed stage microscope |
| Option: |
Motorized XY stage, High-speed Z stage, VAAS |
| Software |
|
| Display/image generation: |
2D analysis, 3D volume rendering/orthogonal, 4D analysis, spectral unmixing |
| Image format: |
JP2, JPG, TIFF, BMP, GIF, PNG, ND2, JFF, JTF, AVI, ICS/IDS |
| Application: |
FRAP, FLIP, FRET, photo activation, three-dimensional time-lapse imaging, multipoint time-lapse imaging, colocalization |
| Control computer |
|
| OS: |
Microsoft Windows® XP 32bit SP (English version) |
| CPU: |
Intel Xeon 5160 (3GHz/1333MHz/dual core) or higher |
| Memory: |
4GB or more |
| Hard disk: |
SAS (15,000rpm), 160GB or more x2, RAID 0 configuration |
| Data transfer: |
Dedicated data transfer I/F |
| Monitor: |
1600 x 1200 or higher resolution, 2 LCD monitor configuration recommended |
| Installation condition: |
Temperature 5 - 35°C, humidity 65% (RH) or less (non-condensing) |
| Power Source |
|
| Controller: |
Input voltage: 100–240VAC ±10% 50–60Hz Current rating: 5A @100VAC Overcurrent protection: main breaker 15A |
| LU-LR 4-laser power source rack: |
Power source for Ar laser and control circuit: 100VAC, 15A/115VAC, 15A/230VAC, 7.5A, 50/60Hz (breaker 15A) Power source for lasers except Ar laser: 100VAC, 3A/115VAC, 3A/230VAC, 1.5A, 50/60Hz (breaker 5A) |